Performance Data

  • Superior measuring ranges: 5 - 400 mg/L (whole blood); 2-160 mg/L (plasma)
  • High precision: plasma CV of 2.4% (5.7 mg/dL); whole blood CV of 4.1% (3.96 mg/dL)
  • Prozone security up to 1800 mg/L
  • No significant interferences
  • Calibration stability: 9 months (from date of calibration)
  • Shelf life: 18 months (from date of production)

Precision in Whole Blood

Total precision
n = 20
Mean
[mg/L]
SD
[mg/L]
CV
[%]
Sample 14.330.13 3.1    
Sample 210.60.22 2.1
Sample 341.50.69 1.7

Method Comparison

A) DiaSys CRP IS using whole blood shows excellent correlation to lab analyzer

B) Competitor system shows higher deviations/outliers compared to lab analyzer