Performance Data
- Superior measuring ranges: 5 - 400 mg/L (whole blood); 2-160 mg/L (plasma)
- High precision
- Prozone security up to 1800 mg/L
- No significant interferences
- Calibration stability: 9 months (from date of calibration)
- Shelf life: 18 months (from date of production)
Precision in Whole Blood
Total precision n = 20 | Mean [mg/L] | SD [mg/L] | CV [%] |
---|---|---|---|
Sample 1 | 4.32 | 0.144 | 3.33 |
Sample 2 | 10.4 | 0.388 | 3.71 |
Sample 3 | 41.6 | 1.01 | 2.43 |